Use / program section: | Ventilation of safety cabinets |
Width: | 250 mm |
Depth: | 100 mm |
Height: | 160 mm |
Weight: | 5 kg |
dim. 210x210x215 mm (WxDxH) ATEX-declaration of conformity in accordance with directive 94/9/EC and type certificate Consisting of: Cage made of sheet steel in silver with sparkling effect "Integrated, modern LED-Display ""Blue Mode""." Plug-in access to the 230V / 50 Hz power supply via IEC-60320 C13 appliance-side connector AC ventilator integrated Optical and acoustic warning in case of errors. Two floating contacts (24V/2A) for signal trans- mission to an external control room. Functions: Monitoring of the air flow and power supply. Internal storage accumulator supplies the monitoring systems for 4 hours in the event of a power failure, display with basic error scale. Last 50 alarm messages will be saved including date, time, type of error and identification of staff member.